Shandon Cytospin 4 Standard Workflow Diagram

SPECIMEN EXAMINATION
Page 89

    • Specimen origin – precise anatomical size
    • Volume of specimen
    • Physical appearance of specimen – colour, viscosity, whether it is homogenous or contains tissue fragments or blood

DETERMINE THE CELL COUNT
Page 90
The concentration chosen should allow enough space for cells to form a monolayer with minimum overlap, but not leave too much space between cells.
Average cells (diameter 10-12 microns) produce excellent Cytospin preparations at cell densities of 1×106 cells per ml. Larger cells will require lower concentrations; smaller cells, cell organelles or bacteria will require higher concentrations.

CONCENTRATE OR DILUTE THE SPECIMEN AS REQUIRED
Page 94 (to concentrate);
Page 95 (to dilute)

LOAD SHANDON CYTOSPIN SAMPLE CHAMBERS
Page 40-44 and page 95
Place the EZ Cytofunnels into the sealed head. Make sure they are distributed evenly so that the Shandon Cytospin 4 is not out of balance.
Load the EZ Cytofunnels after they have been inserted into the sealed head.
Do not place more than 0.5ml of a sample in an EZ Cytofunnel sample chamber with white filter card (0.4ml for an EZ Cytofunnel sample chamber with brown filter card or 6ml maximum in an EZ Megafunnel). Make sure that the sample is deposited directly into the bottom of the sample chamber – do not allow the sample to drip down the sides of the chamber.

BE AWARE OF SAMPLES USED. THEY MAY POSE A BIO HAZARD.

SELECT THE TIME, SPEED AND ACCELERATION FOR THE SHANDON CYTOSPIN PROGRAM
Page 44-47 and page 96
Average cells will require an approximate speed of 1000 rpm with medium acceleration.
Large or fragile cells should be spun at a slower speed (e.g. 500-800 rpm) with low acceleration; small cells or bacteria may require higher speeds (e.g. up to 2000 rpm) with high acceleration. Note that the maximum speed for the EZ Megafunnel is 1500 rpm, and the maximum run duration for EZ Single and Double Cytofunnels is 60 minutes.

RUN THE SHANDON CYTOSPIN
Page 48
When the Shandon Cytospin 4 is programmed and the sealed head is loaded, press START to start the run.

UNLOAD CYTOSPIN SAMPLE CHAMBERS
Page 97
The Shandon Cytospin 4 lid will unlock automatically as soon as the Shandon Cytospin has stopped spinning.
As soon as possible after the Shandon Cytospin has stopped spinning, remove the sealed head from the instrument and open it in a biological safety cabinet. Open the sealed head lid and remove the EZ Cytofunnel assemblies.

FIXATION
Page 98
Fix the samples as soon as possible to avoid autolysis.

STAINING
The specimen can now be stained and examined according to laboratory procedures and examined microscopically.

MICROSCOPY AND DIAGNOSIS

The page and paragraph references refer to the Shandon Cytospin 4 Operator Guide A78310250 issue 4

Reading Luciferase and Renilla signal using Promega Dual Luciferase Reporter Assay

Icon

NOTE: in order to read luciferase using this system, cells must be previously transfected with a luciferase reporter construct (either purchased or cloned for gene of interest) and Renilla plasmid (in lab stock)



• Reagents:

Dual Luciferase Reporter Assay System (Promega cat # E1910)

Eppendorfs

Luminometer

Sample plate(s)

PBS

Distilled water

Cell scrapers


• Preparation:

  1. Thaw frozen buffers in 37 C waterbath.
  2. Turn on luminometer to warm up before taking readings.
  3. Dilute Passive Lysis Buffer (PLB) 1:5 in distilled water and mix well.
  4. Prepare Luciferase Assay Reagent II (LARII) by adding 10 ml Luciferase Assay Buffer II to lyophilized substrate. Aliquot LARII in eppenndorfs and store unused aliquots at -80 (stable for up to 1 yr) protected from light.
  5. Prepare an adequate volume of Stop & Glo Reagent to perform desired number of assays. Dilute Stop & Glo Substrate 1:50 with Stop & Glo Buffer in a 15 ml falcon tube.
    100 µl Stop & Glo Reagent/assay * _ assays (aka readings) = __ µl Stop & Glo needed + __ µl pipet error = ____ µl total volume of Stop & Glo Reagent needed
  6. Predispense 100 µl of LARII into appropriate number of eppendorfs to complete desired number of luciferase and Renilla readings.
  7. Check that luminometer is programmed to perform a 3 second premeasurement delay, followed by a 10 second measurement period for each assay.

    Icon

    NOTE: these are default settings on the Kuperwasser luminometer).


• Protocol:

  1. Aspirate media from cells and wash with PBS.
  2. Aspirate PBS and add appropriate amount of diluted PLB (see chart below) to plates (lysis).
  3. Lyse cells using a cell scraper.
  4. Transfer 20 µl of lysis sample to eppenndorf tube containing LARII, mix up and down with a pipet 2-3 times. Place tube in luminometer and take reading.
  5. If luminometer does not print, record the luciferase activity measurement by hand.
  6. Remove eppendorf from machine and add 100 µl Stop & Glo Reagent.
  7. Vortex sample for a few seconds.
  8. Place in luminometer again and take Renilla reading. Record by hand if luminometer does not print.
  9. Discard this tube, and repeat again with next sample.
  10. Calculate Luciferase/Renilla signal intensity, and normalize signal to that from cells transfected with a control-luciferase reporter plasmid and Renilla.

Multiwell Plate

1X PLB

6-well culture plate 500µl
12-well culture plate 250µl
24-well culture plate 100µl
48-well culture plate 65µl
96-well culture plate 20µl

Instructions for the care and use of the Multisizer 3 (PJ Keller)

Multisizer Operation (2010) [PDF]

The Multisizer is intended for quantifying either cells or spheres. We are in possession of two apertures: one with a 100-µm diameter opening and the other with a 560-µm diameter opening. The default on the Multisizer is the 560-µm aperture which is used for counting spheres. If you need to use the 100 µm aperture for counting cells you need to change two things: first the diluent should be changed to regular Isoton II and secondly, the aperture will need to recalibrated for the new diluent and aperture size. Please see section 4.4 in the operator’s manual for changing the aperture tube and section 5 for calibration of the new aperture.

All of the following instructions assume the use of the 560-µm aperture.

Things to consider before starting a run on the Multisizer 3

Do you have enough diluent?

Life would be easy if we just had to use Isoton II as the diluent but because we are using the large aperture we use the 6:4 Isoton II:Glycerol diluent. This is to reduce the noise in the diluent due to the high rate of intake flow that would occur with just the Isoton II alone. For the same reason, this will also help to keep the spheres from breaking apart while going through the aperture by slowing them down.

Note that if you run 6 collections for each sample as indicated below you will need about 50 ml of diluent for each sample. I would make sure there is at least 1 liter of diluent for each 12 or so samples you plan to run to be safe.

To make diluent (2L):

Mix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink).

Filter to remove small particulates that will cause noise in the diluent, I usually use a 0.2 µm bottle-top filter (you can re-use it until the whole 2L is filtered), or we also have circular filters in the Multisizer accessories drawer that will work, though you will have to apply vacuum pressure to use these as well (the solution is too viscous to go through on its own).

Transfer filtered diluent to a vacuum flask if not already in one and apply vacuum pressure to de-gas the diluent (you should see millions of tiny bubbles rise to the surface). This will take 20-60 minutes (or it can go longer).

The diluent can be left in the vacuum flask or transferred to clean bottles by gently pouring to minimize more bubbles. It is best if the diluent is allowed to settle overnight before use to allow any bubbles to settle.

Are you spheres too big?

This should not be a concern if you are working with primary cells but the tumorspheres can get quite large. The aperture is really only accurate at reading between a range of 2-60% of its size, about 11 mm to 336 mm. Large spheres can also clog the aperture, which would be very bad.

Several potential ways to address this:

1. Verify that you large spheres are okay for the multisizer.

I would take a picture of some of the big ones and then use the “Add measurement” feature in the Spot software (EDIT menu) to check the size of these spheres.

2. Alter the conditions of your experiment or run the tumorspheres earlier so they don’t grow so large.

Abnormally large spheres in my experience are more likely to result from initial aggregation so plating at a lower density may help with this. Monitor your spheres daily and count them on a day when the size is manageable (older spheres may just keep getting bigger).

3. Try to filter out the large spheres.

We don’t have a suitable basket filter or other type of filter that would do this currently (our largest is 100 µm) but I know that larger ones do exist and could be found with some research.

Are your spheres too concentrated?

Again, this is likely going to be a problem only for the Tumorspheres. Note that in the panel to the left when you run a sample there will be an indicator of the concentration of the events. The multisizer seems to like the concentration between 0-10%, anything higher and you get a red bar. In Chris’ experience, when she had too high of a concentration, the machine would behave strangely, counting many, many thousands of spheres and not showing a decay with dilution of the sample. It worked better when she diluted the samples or started with less concentrated samples. We don’t know why this is exactly but if you notice that you are running samples at a high >10% concentration then I would suggest that you dilute your sample or plate cells at a lower density to start with.

What is the size cutoff that counts as a ‘sphere’? What trend can I expect?

1. I strongly recommend that you take some pictures of your spheres and do some measurements on them even if you don’t think you will have size issue as above. This will give you an idea of where you want to put your cutoff when doing data analysis (this only needs to be done once). Try to get a field with small and large spheres. Determine what ‘counts’ as a sphere in your assay, we should be consistent between users so those using the same cell line should consult each other. Also it is a good idea to take some of your spheres and take pictures of them in the diluent for comparison. I have found that my primary spheres shrink considerably in the viscous diluent-about 15 mm in diameter on average.

Some examples from my primary spheres:

Single cells are about 5 mm in diameter in media

Enlarged cells and doublets (entosis victims?) are about 15-20 mm in diameter in media

Multicellular ‘spheres’ are about 35-40+ µm in diameter, with most around 50 µm in media

Multicellular ‘spheres’ are about 20-25+ µm in diameter in diluent, I usually set my counting cutoff at 25-30 mm during data analysis; the consensus for tumorsphere users is about 40 mm for cutoff.

2. If you have different conditions, I would look carefully at your wells/plates before running the samples to see what the expected trend should be (sample A has more spheres than sample B). I sometimes will do a hand count of one well to estimate this trend for my own piece of mind. The multisizer count will not be exactly the same as that by hand for several reasons (human error, the entire sample is not counted by the multisizer, sample loss during preparation etc.) but the trend should be the same. The multisizer has pleasantly always replicated the trend I see and is usually very consistent between replicates.

Lastly, remember that spheres run through the multisizer are gone forever so any additional analysis you need to do on the spheres (staining, dissociation for secondary spheres etc.) will need to be done with spheres you set up in parallel to the ones you use for the multisizer. I like to run my samples in triplicate so plan accordingly.

OK, now for operation of the machine! Note: it can be a bit fussy…it helps if it is properly warmed up and you pipet slowly and methodically.

Operation of the Multisizer 3

1. Exchanging Coulter Cleanz storage solution for diluent; preparing for run

  • Turn on the Multisizer with the power switch on the Right side, allow it to warm up for 20 minutes or so. I usually use this time to fill multiple coulter counter buckets with diluent. Fill as many as needed for your samples plus extras with 20 ml, pipetting carefully to avoid bubbles. Cover with caps if it will be a long time before you actually get to your samples to limit dust.
  • Open the software on the desktop (Multisizer 3), make sure the check box with “Connect to Multisizer 3” is checked, then click OK.
  • Loosen the connection of the Coulter Cleanz bucket (front) on the left side of the machine, pull the tube out from the machine and leave it disconnected.
  • Go to SYSTEM→DRAIN SYSTEM, click OK on the text box that reminds you to disconnect the tube.
  • After the system is drained, unscrew the cap from the bottle of coulter cleanz, hose off the wand briefly with dH20 and transfer the cap to the bottle of diluent that is stored to the left of the machine. Make sure to add more diluent slowly if the level looks low. Tighten the cap and reconnect the tube to the side of the machine, push in firmly and re-screw in snugly.
  • Exchange the coulter bucket of coulter cleanz on the stage for a bucket containing diluent. Press the button on the bottom of the front of the stage to raise and lower it.
  • Got to SYSTEM→FILL SYSTEM, the blue coulter cleanz should be exchanged for the clear diluent.
  • Add more diluent to the bucket on the stage (always raise and lower the stage to add diluent to avoid bumping the aperture or the electrode, pipet against the side of the bucket to reduce bubbles). Repeat step g.
  • Add more diluent to the bucket, flush the aperture tube to get rid of bubbles in the aperture: SYSTEM→FLUSH APERTURE TUBE
  • Exchange the bucket on the stage for a fresh bucket of diluent, place used buckets in the sink to avoid confusion.
  • Check the internal waste tank indicator on the bottom left of the screen, if the blue is far to the right or the red is indicted, empty the internal tank: SYSTEM→EMPTY WASTE TANK. This will take a few minutes. Note the machine will not run if the internal tank is full. If the WASTE TANK FULL error message comes up, check this indicator and also the external tank and empty the appropriate one.
  • Check the level in the external waste tank, if this is near full, disconnect it and empty it into the sink, then replace it.

2. Setting up file storage and naming; diluent quality control I

  • Designate a place for your files to save. Use the shortcut folder on the desktop to get to the multisizer samples file folder. Make yourself a folder if you don’t already have one and make a subfolder within this folder for this experiment.
  • The left panel on the software should say READY at the top and have two boxes: one that says Sample Information and one that says SOM.
  • Click CHANGE in the SOM box on the left (or go to SETTINGS→Change SOM).
  • In the ‘control mode’ box, make sure TIME is clicked and set to 3 seconds (our default). The number of runs should be set to 1. The two check boxes should be unchecked.
  • In the second ‘box’ click the THRESHOLD button to check the background noise of the diluent. Click MEASURE NOISE LEVEL. The level should be somewhere less than 14 µm if the aperture is okay and the diluent is okay. If not, ask for help. Click OK. I usually set the Sizing and Counting Threshold at 14 µm (this is where it starts recording data), only because it is consistent with my other runs. You should set this wherever you want but I suggest you set it below the cutoff you will use in data analysis.
  • In the box that starts ‘After Each Run’ the only box that should be checked is ‘save file’ and ‘include pulse data’. Click the DIRECTORY button to tell the machine where to save your files. Browse for the folder you made in ‘b’ and select it.
  • Go back to the very top of the Standard Operating Method window and click OK.
  • In the ‘Sample Information’ box on the top left panel, click CHANGE. This should pop up a new window. Enter in the sample name in the Group ID box (i.e. SUM149, control etc.) and enter in the replicate number in the Sample ID box (i.e. 1, A etc.). Click OK. The current file naming system is set up to name files as GroupID_SampleID_Date_run number.#M3. When collecting 6 runs for each replicate the samples will be named SUM149_A_10 Dec 2008_01.#M3, SUM149_A_10 Dec 2008_02.#M3 etc.

3. Diluent Quality Control II

  • It is imperative that the diluent be reading “quietly” when the diluent is run by itself. Change the sample name to something like ‘background’ so this data is distinguished from your sphere data.
  • Make sure there is at least 20 ml of diluent in the coulter bucket. From the top menu panel click the green START arrow. You should see a graph pop up with red bars indicating counts. There is also a small window that plots counts in black lines on the left panel. You should get somewhere in the range of 0-15 counts if the diluent is behaving properly. Note that sometimes it takes several runs for the diluent to settle down. I think there is sometimes bubbles in the aperture that make it have higher counts.
  • Click start again to do a second run. Refill the bucket with 12.5 ml of diluent by lowering the stage, pipeting diluent into the bucket along the wall of the bucket, and raising the stage. Click start again to do another run. Repeat this process to see if the counts settle in to the range of 0-15, refilling the bucket after every two runs.
  • If the counts remain high there are a few options to try.
    • Make sure that you are not introducing bubbles when pipetting, watch as you aspirate into the pipet and pipet out into the bucket for excessive bubbling…done slowly and carefully this should not be a problem.
    • Try changing to a new bucket of diluent, run again.
    • When the bucket is full, try flushing the aperture tube (SYSTEM→FLUSH APERTURE TUBE), this will aspirate into the aperture and then flush into the bucket. Change to a new bucket of diluent and run again at least twice.
    • If none of the above works, try a new bottle of diluent. Old bottles sometimes will start to have higher counts if there is contamination.
  • Note that if you ever do run out of diluent during a run and introduce bubbles you must refill the bucket and then flush the aperture tube or it won’t run. Change to a fresh bucket after every aperture flush (it just spits bubbles into the diluent).
  • Leave the aperture immersed in the diluent and go prepare samples.

A NOTE on the waste tanks. There is an internal waste tank that is monitored by the blue bar at the bottom left of the screen. This empties to the external waste tank when you go to SYSTEM→EMPTY WASTE TANK. Make sure you monitor the levels in the waste tank and empty it when it gets near full (a red bar appearing means you are full). If you do not do this and the tank becomes full during a run it will stop everything and give you a warning message. It will often not record the data from that run even though it has sucked up the sample. This usually makes me pretty cranky when this happens so prevention is the best cure! I keep an eye on it and check it before I start a sample. It helps to empty it before you run any samples after you have filled the system as this process seems to fill up the waste tank. I usually empty the external waste tank into the sink when it gets near full or when I am done with the machine, run plenty of water to flush.

4. Sample Preparation

The method of preparation will depend on your preference, as long as you are consistent with how you run your data, it shouldn’t really matter. The following instructions are the way that I run my data. I typically will prepare only 6-12 samples at time to avoid having some samples sit for extended periods of time.

  • Collect spheres to a 15 ml conical tube, wash out the well with PBS and add this to the conical tube.
  • Spin down spheres at 6-800 rpm for 4-5 min. Aspirate off the media and resuspend spheres in about 1 ml of media. Note I have found that my spheres will disintegrate after extended incubation in PBS so I recommend you resuspend in media.
  • Bring samples and a tube of media for a control over to the multisizer, where you should have prepared enough buckets with 20 ml of diluent for your samples plus a few extra.

5. Running your samples

  • Change the sample name to Media control (or something like that). To change the sample name after a run, click the RESET button on the left panel near the bottom. This will bring you back to the panel where you see the Sample Information box, click CHANGE, and type in your new sample name.
  • With a 1 ml pipet or a p1000, pipet 1 ml of media into the 20 ml of diluent in a bucket, inserting the pipet tip into the diluent so the sample mixes well.
  • Place the bucket on the stage. Do 2 runs, add 12.5 ml diluent, do 2 more runs, add 12.5 ml diluent. This is 4 runs total; for the media control, you should see a low number of counts (50-60 max in any one run). I save these runs and use them to subtract out background from my samples during data analysis.
  • Periodically change to a fresh bucket of diluent and run a blank sample. You can change the sample name for this if you want. I usually just note that the 5th run for any sample is my blank. The counts should be in the 0-15 range again. I usually just add more diluent to this bucket and re-use it as my blank between runs but if you notice the counts creeping up, change to a fresh bucket.
  • Proceed with your samples, repeating steps a-c for each, running a blank is optional between samples. I have found that it is nearly always 0-15 counts and therefore it is a waste of diluent to run a blank between each sample but periodically checking the diluent performance may be warranted if your samples seem to be behaving oddly.
    • Note: you should see decay in the counts with your subsequent runs. If you don’t, something is wrong (unfortunately, for some cell lines, it behaves this way and I don’t know why). I recommend you do all 4 runs for a few samples to see if it is behaving properly before resorting to alternate methods. I typically see the counts behave with the following pattern (give or take; note schematic is from when I used to do 6 runs per sample, I have found runs 5 and 6 have a lot of noise in them so have stopped at 4):
    • The media will sometimes have a lot of counts in runs 5&6 or it won’t decay, I generally take these as bogus and use the first 4 runs to estimate a background subtraction factor (it is an estimate after all).
    • Note that you end up reading about 90% of your input by doing four to six runs. More runs will give you more counts but it starts getting to a point where it is just noise and is not worth the time. Thus, it will still be an estimate of the total number of spheres counted (though a hand count will also be an estimate, in some respects).
    • If you start seeing a lot of counts in the blanks, try some of the tips in part 3d.

6. Exchanging the diluent for Coulter Cleanz; shutting down the machine

Repeat the procedure in step 1, except you are now exchanging the diluent for the Coulter Cleanz. There should be a small bottle of Coulter Cleanz that you can use to fill the bucket between FILL SYSTEM steps. More Coulter Cleanz is stored on the shelf to the right of the Multisizer. Please store upright as these leak!

7. Data retrieval and analysis (current method, I think there is a way to set the files to save as converted in the first place but I haven’t tested it yet)

  • Before you begin, make a copy of your data folder and use the copy to manipulate the files (in case you need to go back to the originals)
  • Overlay the 4 files from each sample: Go to FILE→OVERLAY and select the files to overlay (ex. Sample A 1-4). Click ADD, then OK. A graph should appear with 4 sets of lines corresponding to your 4 runs. Note you can do this with more than 4 samples since this is just converting pulses to size and not adding runs. I will often open up as many as it will let me.
  • In the graph window go to ANALYZE→Convert pulses to size. In the new window:
    • Change Log Diameter to Lin Diameter
    • Check the box to do all runs
    • Make 300 bins between 14 µm (or wherever you started collecting data) and 336 µm (this is roughly 2-60%, the accurate range of the aperture). If you want 300 bins with 1 µm increments use 20-320 as your range.
    • Coincidence correction should be on, Multisizer II edit off
  • Click OK, the lines in the graph should convert to steps.
  • In the graph window, go to RUN FILE→Save All.
  • Back in the main Multisizer window go to FILE→Add runs. Select the 4 runs from one of your samples that you just converted. Click ADD, then OK. You should now see one red step graph that represents your added runs.
  • In the graph window, got to RUN FILE→Save As. Give this file a new name to distinguish it from the individual files.
  • From RUN FILEàExport Data. Check the Listing box (leave the others as is). This will now export the data as a CSV file that can be opened in Excel.
  • In Excel, open the CSV file and you should see data as follows:
    | Bin Number | Bin Diameter | Diff. | Diff. |
    (Lower) Number Volume
    um %
    1 14 88 1.12643
    2 15.0733 88 1.39408
    3 16.1467 80 1.54648
    4 17.22 79 1.84059

The bin number refers to the 300 bins you made. Bin diameter (lower) is the lower size limit of the bin. So in Line 1, 88 (Diff. Number) refers to the number of counts between 14 µm (the lower limit of the bin) and 15.0733 µm (the lower limit of the next bin). I disregard Diff. Volume %.

To get an estimate of the total counts, sum the numbers in the Diff. Number column greater than your size cutoff of choice (ex. 40 mm). I usually will do the same for the media control and subtract this out as background.

Instructions for the care and use of the Multisizer 3

Multisizer Operation [PDF]

The Multisizer is intended for quantifying either cells or spheres. We are in possession of two apertures: one with a 100-µm diameter opening and the other with a 560-µm diameter opening. The default on the Multisizer is the 560-µm aperture which is used for counting spheres. If you need to use the 100 µm aperture for counting cells you need to change two things: first the diluent should be changed to regular Isoton II and secondly, the aperture will need to recalibrated for the new diluent and aperture size. Please see section 4.4 in the operator’s manual for changing the aperture tube and section 5 for calibration of the new aperture.

All of the following instructions assume the use of the 560-µm aperture.

Things to consider before starting a run on the Multisizer 3

Do you have enough diluent?

Life would be easy if we just had to use Isoton II as the diluent but because we are using the large aperture we use the 6:4 Isoton II:Glycerol diluent. This is to reduce the noise in the diluent due to the high rate of intake flow that would occur with just the Isoton II alone. For the same reason, this will also help to keep the spheres from breaking apart while going through the aperture by slowing them down.

Icon

If you run 6 collections for each sample as indicated below you will need about 50 ml of diluent for each sample. I would make sure there is at least 1 liter of diluent for each 12 or so samples you plan to run to be safe.

To make diluent (2L):

Mix 920 ml of 87% glycerol (on shelf) with 1080 ml of Isoton II (in cabinet), stir in a flask until homogenized (the glycerol will sink).

Filter to remove small particulates that will cause noise in the diluent, I usually use a 0.2 µm bottle-top filter (you can re-use it until the whole 2L is filtered), or we also have circular filters in the Multisizer accessories drawer that will work, though you will have to apply vacuum pressure to use these as well (the solution is too viscous to go through on its own).

Transfer filtered diluent to a vacuum flask if not already in one and apply vacuum pressure to de-gas the diluent (you should see millions of tiny bubbles rise to the surface). This will take 20-60 minutes (or it can go longer).

The diluent can be left in the vacuum flask or transferred to clean bottles by gently pouring to minimize more bubbles. It is best if the diluent is allowed to settle overnight before use to allow any bubbles to settle.

Are your spheres too big?

This should not be a concern if you are working with primary cells but the tumorspheres can get quite large. The aperture is really only accurate at reading between a range of 2-60% of its size, about 11 µm to 336 µm. Large spheres can also clog the aperture, which would be very bad.

Several potential ways to address this:

  1. Verify that your large spheres are okay for the multisizer.
    I would take a picture of some of the big ones and then use the “Add measurement” feature in the Spot software (EDIT menu) to check the size of these spheres.
  2. Alter the conditions of your experiment or run the tumorspheres earlier so they don’t grow so large.
    Abnormally large spheres in my experience are more likely to result from initial aggregation so plating at a lower density may help with this. Monitor your spheres daily and count them on a day when the size is manageable (older spheres may just keep getting bigger).
  3. Try to filter out the large spheres. We don’t have a suitable basket filter or other type of filter that would do this currently (our largest is 100 µm) but I know that larger ones do exist and could be found with some research.

What is the size cutoff that counts as a ‘sphere’? What trend can I expect?

1. I strongly recommend that you take some pictures of your spheres and do some measurements on them even if you don’t think you will have size issue as above. This will give you an idea of where you want to put your cutoff when doing data analysis (this only needs to be done once). Try to get a field with small and large spheres. Determine what ‘counts’ as a sphere in your assay, we should be consistent between users so those using the same cell line should consult each other. Also it is a good idea to take some of your spheres and take pictures of them in the diluent for comparison. I have found that my primary spheres shrink considerably in the viscous diluent-about 15 µm in diameter on average.

Some examples from my primary spheres:

Single cells are about 5 µm in diameter in media
Enlarged cells and doublets (entosis victims?) are about 15-20 µm in diameter in media
Multicellular ‘spheres’ are about 35-40+ µm in diameter, with most around 50 µm in media
Multicellular ‘spheres’ are about 20-25+ µm in diameter in diluent, I usually set my counting cutoff at 25-30 µm during data analysis; the consensus for tumorsphere users is about 40 µm for cutoff.

2. If you have different conditions, I would look carefully at your wells/plates before running the samples to see what the expected trend should be (sample A has more spheres than sample B). I sometimes will do a hand count of one well to estimate this trend for my own piece of mind. The multisizer count will not be exactly the same as that by hand for several reasons (human error, the entire sample is not counted by the multisizer, sample loss during preparation etc.) but the trend should be the same. The multisizer has pleasantly always replicated the trend I see and is usually very consistent between replicates.

Lastly, remember that spheres run through the multisizer are gone forever so any additional analysis you need to do on the spheres (staining, dissociation for secondary spheres etc.) will need to be done with spheres you set up in parallel to the ones you use for the multisizer. I like to run my samples in triplicate so plan accordingly.

OK, now for operation of the machine! Note: it can be a bit fussy…it helps if it is properly warmed up and you pipet slowly and methodically.

Operation of the Multisizer 3

  1. Exchanging Coulter Cleanz storage solution for diluent; preparing for run
    1. Turn on the Multisizer with the power switch on the Right side, allow it to warm up for 20 minutes or so. I usually use this time to fill multiple coulter counter buckets with diluent. Fill as many as needed for your samples plus extras with 20 ml, pipetting carefully to avoid bubbles. Cover with caps if it will be a long time before you actually get to your samples to limit dust.
    2. Open the software on the desktop (Multisizer 3), make sure the check box with “Connect to Multisizer 3” is checked, then click OK.
    3. Loosen the connection of the Coulter Cleanz bucket (front) on the left side of the machine, pull the tube out from the machine and leave it disconnected.
    4. Go to SYSTEM→DRAIN SYSTEM, click OK on the text box that reminds you to disconnect the tube.
    5. After the system is drained, unscrew the cap from the bottle of coulter cleanz, hose off the wand briefly with dH20 and transfer the cap to the bottle of diluent that is stored to the left of the machine. Make sure to add more diluent slowly if the level looks low. Tighten the cap and reconnect the tube to the side of the machine, push in firmly and re-screw in snugly.
    6. Exchange the coulter bucket of coulter cleanz on the stage for a bucket containing diluent. Press the button on the bottom of the front of the stage to raise and lower it.
    7. Go to SYSTEM→FILL SYSTEM, the blue coulter cleanz should be exchanged for the clear diluent.
    8. Add more diluent to the bucket on the stage (always raise and lower the stage to add diluent to avoid bumping the aperture or the electrode, pipet against the side of the bucket to reduce bubbles). Repeat step g.
    9. Add more diluent to the bucket, flush the aperture tube to get rid of bubbles in the aperture: SYSTEM→FLUSH APERTURE TUBE
    10. Exchange the bucket on the stage for a fresh bucket of diluent, place used buckets in the sink to avoid confusion.
    11. Check the internal waste tank indicator on the bottom left of the screen, if the blue is far to the right or the red is indicated, empty the internal tank: SYSTEM→EMPTY WASTE TANK. This will take a few minutes. Note the machine will not run if the internal tank is full. If the WASTE TANK FULL error message comes up, check this indicator and also the external tank and empty the appropriate one.
    12. Check the level in the external waste tank, if this is near full, disconnect it and empty it into the sink, then replace it.
  2. Setting up file storage and naming; Diluent Quality control I
    1. Designate a place for your files to save. Use the shortcut folder on the desktop to get to the multisizer samples file folder. Make yourself a folder if you don’t already have one and make a subfolder within this folder for this experiment.
    2. The left panel on the software should say READY at the top and have two boxes: one that says Sample Information and one that says SOM.
    3. Click CHANGE in the SOM box on the left (or go to SETTINGS→Change SOM).
    4. In the ‘control mode’ box, make sure TIME is clicked and set to 3 seconds (our default). The number of runs should be set to 1. The two check boxes should be unchecked.
    5. In the second ‘box’ click the THRESHOLD button to check the background noise of the diluent. Click MEASURE NOISE LEVEL. The level should be somewhere less than 14 µm if the aperture is okay and the diluent is okay. If not, ask for help. Click OK. I usually set the Sizing Threshold at 14 µm (this is where it starts recording data), only because it is consistent with my other runs. You should set this wherever you want but I suggest you set it below the cutoff you will use in data analysis.
    6. In the box that starts ‘After Each Run’ the only box that should be checked is ‘save file’ and ‘include pulse data’. Click the DIRECTORY button to tell the machine where to save your files. Browse for the folder you made in ‘b’ and select it.
    7. Go back to the very top of the Standard Operating Method window and click OK.
    8. In the ‘Sample Information’ box on the top left panel, click CHANGE. This should pop up a new window. Enter in the sample name in the Group ID box (i.e. SUM149, control etc.) and enter in the replicate number in the Sample ID box (i.e. 1, A etc.). Click OK. The current file naming system is set up to name files as GroupID_SampleID_Date_run number.#M3. When collecting 6 runs for each replicate the samples will be named SUM149_A_10 Dec 2008_01.#M3, SUM149_A_10 Dec 2008_02.#M3 etc.
  3. Diluent Quality Control II
    1. It is imperative that the diluent be reading “quietly” when the diluent is run by itself. Change the sample name to something like ‘background’ so this data is distinguished from your sphere data.
    2. Make sure there is at least 20 ml of diluent in the coulter bucket. From the top menu click the green START arrow or click START from the bottom of the left panel. You should see a graph pop up with red bars indicating counts. There is also a small window that plots counts in black lines on the left panel. You should get somewhere in the range of 0-15 counts if the diluent is behaving properly. Note that sometimes it takes several runs for the diluent to settle down. I think there are sometimes bubbles in the aperture that make it count a lot.
    3. Click start again to do a second run. Refill the bucket with 12.5 ml of diluent by lowering the stage, pipeting diluent into the bucket along the wall of the bucket, and raising the stage. Click start again to do another run. Repeat this process to see if the counts settle in to the range of 0-15, refilling the bucket after every two runs.
    4. If the counts remain high there are a few options to try.
      1. Make sure that you are not introducing bubbles when pipetting, watch as you aspirate into the pipet and pipet out into the bucket for excessive bubbling…done slowly and carefully this should not be a problem.
      2. Try changing to a new bucket of diluent, run again.
      3. When the bucket is full, try flushing the aperture tube (SYSTEM→FLUSH APERTURE TUBE), this will aspirate into the aperture and then flush into the bucket. Change to a new bucket of diluent and run again at least twice.
      4. If none of the above works, try a new bottle of diluent. Old bottles sometimes will start to have higher counts.
    5. Note that if you ever do run out of diluent during a run and introduce bubbles you must refill the bucket and then flush the aperture tube or it won’t run. Change to a fresh bucket after every aperture flush (it just spits bubbles into the diluent).
    6. Leave the aperture immersed in the diluent and go prepare samples.
  4. Sample Preparation
    The method of preparation will depend on your preference, as long as you are consistent with how you run your data, it shouldn’t really matter. The following instructions are the way that I run my data. I typically will prepare only 6-12 samples at time to avoid having some samples sit for extended periods of time.

    1. Collect spheres to a 15 ml conical tube, wash out the well with PBS and add this to the conical tube.
    2. Spin down spheres at 6-800 rpm for 4-5 min. Aspirate off the media and resuspend spheres in about 1 ml of media. Note I have found that my spheres will disintegrate after extended incubation in PBS so I recommend you resuspend in media.
    3. Bring samples and a tube of media for a control over to the multisizer, where you should have prepared enough buckets with 20 ml of diluent for you samples plus a few extra.
  5. Running your samples
    1. If the machine has been sitting unattended for a long time, it is a good idea to check that the diluent is still running with 0-15 counts.
    2. Change the sample name to Media control (or something like that). To change the sample name after a run, click the RESET button on the left panel near the bottom. This will bring you back to the panel where you see the Sample Information box, click CHANGE, and type in your new sample name.
    3. With a 1 ml pipet or a p1000, pipet 1 ml of media into the 20 ml of diluent in a bucket, inserting the pipet tip into the diluent so the sample mixes well.
    4. Place the bucket on the stage. Do 2 runs, add 12.5 ml diluent, do 2 more runs, add 12.5 ml diluent, do 2 more runs. This is 6 runs total; for the media control, you should see a low number of counts (50-60 max in any one run). I save these runs and use them to subtract out background from my samples during data analysis.
    5. Change to a fresh bucket of diluent and run a blank sample. You can change the sample name for this if you want. I usually just note that the 7th run for each sample is my blank. The counts should be in the 0-15 range again. I usually just add more diluent to this bucket and re-use it as my blank between runs but if you notice the counts creeping up, change to a fresh bucket.
    6. Proceed with your samples, repeating steps b-e for each.
      1. Note: you should see decay in the counts with your subsequent runs. If you don’t, something is wrong (unfortunately, for some cell lines, it behaves this way and I don’t know why). I recommend you do all 6 runs for a few samples to see if it is behaving properly before resorting to alternate methods. I typically see the counts behave with the following pattern (give or take):
      2. The media will sometimes have a lot of counts in runs 5&6 or it won’t decay, I generally take these as bogus and use the first 4 runs to estimate a background subtraction factor (it is an estimate after all).
      3. Note that you end up reading about 90% of your input by doing six runs. More runs will give you more counts but it starts getting to a point where it is just noise and is not worth the time. Thus, it will still be an estimate of the total number of spheres counted (though a hand count will also be an estimate, in some respects).
      4. If you start seeing a lot of counts in the blanks, try some of the tips in part 3d.
  6. Exchanging the diluent for Coulter Cleanz; shutting down the machine
    Repeat the procedure in step 1, except you are now exchanging the diluent for the Coulter Cleanz. There should be a small bottle of Coulter Cleanz that you can use to fill the bucket between FILL SYSTEM steps. More Coulter Cleanz is stored on the shelf to the right of the Multisizer. Please store upright as these leak!
  7. Data retrieval and analysis (current method, I think there is a way to set the files to save as converted in the first place but I haven’t tested it yet)
    1. Before you begin, make a copy of your data folder and use the copy to manipulate the files (in case you need to go back to the originals)
    2. Open up the 6 files from each sample. Go to FILE→OVERLAY and select the files to overlay (ex. Sample A 1-6). Click ADD, then OK. A graph should appear with 6 sets of lines corresponding to your 6 runs
    3. In the graph window go to ANALYZE→Convert pulses to size. In the new window:
      1. Change Log Diameter to Lin Diameter
      2. Check the box to do all runs
      3. Make 300 bins between 14 µm (or wherever you started collecting data) and 336 µm (this is roughly 2-60%, the accurate range of the aperture).
      4. Coincidence correction should be on, Multisizer II edit off
        Click OK, the lines in the graph should convert to steps.
    4. In the graph window, go to RUN FILE→Save All.
    5. Back in the main Multisizer window go to FILE→Add runs. Select the same runs you just converted. Click ADD, then OK. You should now see one red step graph that represents your added runs.
    6. In the graph window, got to RUN FILE→Save As. Give this file a new name to distinguish it from the individual files.
    7. From RUN FILE→Export Data. Check the Listing box (leave the others as is). This will now export the data as a CSV file that can be opened in Excel.
    8. In Excel, open the CSV file and you should see data as follows:
      Bin Number Bin Diameter
      (Lower) µm
      Diff. Number Diff. Volume
      %
      1 14 88 1.12643
      2 15.0733 88 1.39408
      3 16.1467 80 1.54648
      4 17.22 79 1.84059

The bin number refers to the 300 bins you made. Bin diameter (lower) is the lower size limit of the bin. So in Line 1, 88 (Diff. Number) refers to the number of counts between 14 µm (the lower limit of the bin) and 15.0733 µm (the lower limit of the next bin). I disregard Diff. Volume %.

To get an estimate of the total counts, sum the numbers in the Diff. Number column greater than your size cutoff of choice (ex. 40 mm). I usually will do the same for the media control and subtract this out as background.

FACS Calibure use and hints (PJ Keller)

FACS Calibur use and hints [PDF]

Flow Cytometry on the FACS Calibur: General Instructions

NOTE: The software has been updated since these instructions were written, while they are still good general instructions on using the Caliber, there are some differences in operation of the software so use with caution!

Sign up for time with the calendar in the cytometer room, 8th floor Jaharis

http://cytometry.med.tufts.edu (note you need a password to do this, to get one, contact Allen or Steve)

Allen.Parmelee@tufts.edu or Stephen.Kwok@tufts.edu (for questions)

Use of the cytometer

  • Check that the sheath tank has enough fluid and that it is pressurized
    • To add fluid, vent the pressure with the switch on the right side, add dH20, recap tightly and then repressurize
  • Log in to the computer
    • Login: Kuperwasser
    • Password: 75Kneel@nd
  • Click on the CellQuest Icon
    • From the Acquire menu, select ‘connect to cytometer’
    • From the Cytometer menu, select Detector/Amps and Compensation
    • From the Plots menu, select dot plots or histogram (or both)
      • The plot source should be acquisition unless you want to gate live cells at the cytometer, then, use acquisition to analysis
      • Make a FSC vs. SSC dot plot, histograms for each channel (FL1-FL4) and dot plots for compensation if needed (FL1 vs. FL2, FL2 vs. FL3, FL3 vs. FL4 etc.)
    • From the Acquire menu, select parameter description
      • Make a new folder on the desktop to save your files in
      • Name the file (Set Sample ID as identifier when you hit the file button), change w/ each new sample
      • Label the axes (EpCAM etc), check the detector amps for correlation between P number and channel
  • Hit run-low on the cytometer to run the unstained sample
    • Check the setup box in the acquire box
    • Make sure P1 is thes to E-1 in Detector/Amps for large cells
    • Set FSC and SSC scales to linear, log for everything else (except PI if you don’t expect many dead cells)
    • Adjust all of the compensation sliders to 0% if they are not already there
    • Adjust the voltage for FSC to bring the cells to the middle of the dot plot, do the same for SSC
    • Check the histograms for the different channels, adjust the voltage for that channel so the peak is between 100 and 101 and not too much of the peak is smooshed against the axis
    • Periodically pause, abort and then resume the run so the counts are cleared off the plots so you can see where the peaks fall
  • If you do not need to compensate samples then you are ready to collect data
    • Vortex sample
    • Uncheck ‘setup’ and click acquire
    • Count to 20,000 to 100,000 events (Acquire menu, counter window to change, if you gated live cells change it to count to your number within the gate but makes sure it saves all counts in the data file)
    • Change the file name with each sample
  • Compensation…note this is not a complete description and the jury is still out on whether we want to do this by hand or with software
    • Use single stained samples to adjust the compensation
    • Compensation is a way to account for bleeding of the signals between channels
    • With a dot plot of FL1 vs. FL2 and a single stain of an antibody fluorescing in FL1, adjust the compensation so the cells positive for FL1 do not exceed a value of 101 for FL2. Use the minimum amount of compensation (FL2 = FL2-%FL1) to achieve this.
    • Repeat with additional single stained controls (FL2 vs FL1, adjust FL1 = FL1-%FL2; FL2 vs FL3, adjust FL3 = FL3-%FL2; FL3 vs FL2, adjust FL2 = FL2-%FL3 etc.)
  • Every FACS run should include:
    • Unstained sample
    • Isotype controls for each antibody
    • Single stained samples for setting compensation manually or on the software
    • Stain of interest (double, triple etc.)
    • If doing PI staining, use PI at 0.1 mg/ml and add near time of running the sample (i.e. at the Calibur), run these samples last in your run if possible
      • You may need to run a PI-labeled sample earlier for compensation…if so, run bleach for 30 sec then water for 1’ to get rid of PI before proceeding with non-PI samples
  • When done run bleach for 5 minutes (very important to prevent clogs if PI was used) and then run dH20 for 5’
  • Place the Cytometer on standby when done using it
  • Quit CellQuest, transfer data files to a USB drive, trash the data on the computer (do not store on the computer)
  • Check that the sheath fluid levels, if needed, depressurize, fill the tank and then repressurize for the next user

Channels corresponding to various color fluorophores:

FL1 = FITC, GFP, YFP, Hoechst

FL2 = PE, PI for DNA content

FL3 = PerCP, PE-Cy5, PerCP-Cy5.5, PI for viability

FL4 = APC

Compensation for commonly used ones (limited to Patty’s working knowledge):

FITC and GFP will bleed in FL2

PE bleeds slightly in FL1

PE-Cy5 bleeds a lot in FL4

PerCP-Cy5.5 bleeds minimally in F2 and FL4

APC has minimal bleed in FL3

PI bleeds into both FL2 (a lot) and FL4 (a little)